Ziehm Vision

2007 Ziehm Vision C-Arm

ID
59445
Type
C-Arm
Manufacturer
Ziehm
Model
Vision
Year of MFG
2007
Condition
Functional - Installed

Details

Condition Notes
This 2007 Ziehm Vision is experiencing an issue with the collimator. The collimator can be adjusted, but when an image is taken, the collimator snaps back to its open position. The machine otherwise fluoros and takes static images. See video.
Date of Availability
2004-05-01
Are the manuals present?
no
Are the software disk(s) present?
no
Is the equipment under service?
no
Is the device digital or analog?
Analog
Phantom(s) Included?
no
Egress Description
First floor, average egress. Three or four corners, 200' to exit, no obstacles.

Images

Equipment Listing Images Gallery (16 images)
Click image to zoom in
Previous Review Next Review

Documents

Recent Service Reports
Additional Supporting Documents

Notes

System Options & Presets:

SYSTEM OPTIONS:

Cineruns  (4)

Subtraction (1)

DICOM Storage (1)

DICOM Viewer(1)

GRID regulation (1)

Motion detection (1)

Automatic does reduction (1)

Hard Disc Capacity (5000)

Xray Key Switch enabled (1)

 

Generator power 2KW_STAT

 

PRESETS:

Heart

Uro

Endo

Abdomen

Soft

DSA

MSA

RSA

 

Description of Problems:

This machine does not effectively collimate.  One can set the parameters for collimation, but when the tube is engaged the parameters are automatically reset effectively rendering the collimation useless.

The problem is that the machine software won't marry to the collimator.  

The machine still fluoros and takes static images, it just doesn't collimate.  Please watch the videos and download the service history and physicist reports from this page.  

Ask any question you wish.  You are welcome to inspect in person, but we can also perform a web-based inspection (Skype/Whatsapp/Facetime/etc.) as well.

Motion Test:

We foolishly forgot to test the powered up/down motion of the C-Arm in the video.  It works.  

Machine Test (WATCH!):

Motion Test Video: